INNOVA E3 / E3 with CMS

The INNOVA E3 is designed for use in clinics and hospitals. It has a capacity for two endoscopes, and can be installed under-counter or free standing. An optional base is available, providing an ergonomic working height for the load carrier and space for the chemical bottles. The Innova E3 has a 36-minute cycle time and the bar code scanner allows to record scopes' serial numbers as well as staff identification. This information, together with cycle data is recorded on the server.

This model ensures a flexible and reliable reprocessing process in offering three unique technical solutions:

  • The pressure chamber system
  • The single channel adaptation principle
  • The INNOVA CMS Option (Individual Channel Check)

Standard features include

  • Two-scope capacity
  • Single manual hinged door
  • Single-use chemistry
  • Superior drying system
  • Washing and disinfection in about 36 minutes
  • Endoscope bar coding identification
  • Print-out after every cycle1
  • Thermal disinfection program for the chamber and the entire washing system
  • Internet based remote maintenance with live support (optional)

Technical services

  • Installation by factory trained technicians
  • Preventative maintenance and repair
  • Customized service contracts
  • In-Service training provided by a clinical specialist

If you require a service contract, we can tailor one to meet your explicit needs.
An extensive inventory of parts is also at your disposal.

Windows Embedded INNOVA E3-CMS
STAND-ALONE Model

Specifications

   

INNOVA E3

INNOVA E3-CMS

Dimensions Free-standing unit W/H/D
Under-counter unit W/H/D
Wash chamber W/H/D
Volume
800 x 1440 x 680 mm
800 x 880 x 680 mm
540 x 510 x 535 mm
147 L
1000 x 1440 x 800 mm
1200 x 860 x 680 mm
540 x 510 x 535 mm
147 L
Material Panels and washing chamber: stainless steel AISI 304
Weight   150 kg 175 kg
Electric connection 208V, 3 ph, 60Hz, 16 amp    
Connections Cold, warm, DI water
Waste water
Drain water hose
3/4“ 2 – 5 bar
1 1/2“ 30 l/min
DN 20
Wash Power Fresh water system, pump capacity 420 L/min, tank heating 7 kW
Drying High pressure turbine 150 m3/h, option: heating 3.4 kW
Exhaust air DN 70, drying temperature 45°C (optionally up to 110°C)
Option: exhaust air condenser
DI water boiler Thermal disinfection, volume 20 L, heating 9 kW with heat recovery
Leak Test Device
(independent)
Automatic leak test system to check endoscopes, with analog pressure switch
Dosing 2 dosing systems with dosing supervision and dynamic dosing
Microprocessor Controller 3 programs for flexible endoscopes, 1 program for MIS instruments, 2 programs for thermal disinfection (optional), 6 free program places, RS232 interface for communication with PC, connection possibility for barcode reader and printer interface.


1Optional printer required.